| Adjustable volume dispensers |
|
Inaccuracy (Rel. err.%) |
Imprecision (CV %) |
| Cat. No. |
Vol. mL |
Div. mL |
Min. vol. |
Mid. vol. |
Max. vol. |
Min. vol. |
Mid. vol. |
Max. vol. |
| 521.020 |
2 - 20 |
0.1 |
< ± 1.5% |
< ± 1.1% |
< ± 0.6% |
< 0.5% |
< 0.35% |
< 0.1% |
| 521.050 |
10 - 50 |
1.0 |
< ± 1.5% |
< ± 1.1% |
< ± 0.6% |
< 0.5% |
< 0.35% |
< 0.1% |
| 521.100 |
10 - 100 |
1.0 |
< ± 1.5% |
< ± 1.1% |
< ± 0.6% |
< 0.5% |
< 0.35% |
< 0.1% |
|
Performance values obtained with bi-dist. water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655. |
| Fixed volume dispensers |
| Cat. No. |
Vol. mL |
Inaccuracy (Rel. err.%) |
Imprecision (CV %) |
| 521.F020 |
20 |
< ± 0.6% |
< 0.1% |
| 521.F050 |
50 |
< ± 0.6% |
< 0.1% |
| 521.F100 |
100 |
< ± 0.6% |
< 0.1% |
|
| Specify desired fixed volume within instrument range when ordering. |
|
|
 |